Automated Pressure Calibrator
CPC 6000
62 Mensor/WIKA Operating Instructions - CPC 6000
F$nnnnnnn6X RETURN NULL METER READING NOT SUPPORTED
F$nnnnnnn7X RETURN VACUUM GAUGE READING NOT SUPPORTED
F$nnnnnnn8X Return Clock Reading (Time); $,n ignored
F$nnnnnnn9X Return Pressure Control Limits; $,n ignored
MX Measure Pressure in current pressure units
M$X Measure Pressure in units specified by $
M$nnnnnnnX Measure Pressure in units specified by $; n ignored
M$nnnnnnnnsX Measure Pressure in units specified by $; n,s ignored
Q#X SEQ FUNCTIONS NOT SUPPORTED
RX Not supported
R0X Return to Standard Output Format
R1X Re-initialize DO NOT USE - sets CPC 6000 to command prompt.
R2X RETURN CAL DATA NOT SUPPORTED
R3X Return unit ID string
R5X RETURN QPS TEMPERATURE NOT SUPPORTED
R6X RETURN NULL METER READING NOT SUPPORTED
R7X RETURN VACUUM GAUGE READING NOT SUPPORTED
R8X Return Clock Reading (Time)
R9X Return Pressure Control Limits (Min and Max)
SX Standby Mode
S$X Standby Mode; $ ignored
S$nnnnnnnX Standby Mode; $,n ignored
S$nnnnnnnsX Standby Mode; $,n,s ignored
U$X Change Units to units specified by $
VX Vent Mode in current units
V$X Vent Mode in units specified by $
V$nnnnnnnX Vent Mode in units specified by $; n ignored
V$nnnnnnnsX Vent Mode in units specified by $; n,s ignored
ZX Automatic re-zero