50 Mensor/WIKA Operating Instructions - CPC 6000
Always send commands in one of the following formats:
1. [Command] [Termination];
2. [Command] [Separator] [Data] [Termination];
3. Queries are special instructions in the form:
[Command?] [Termination] where the question
mark, “?”, immediately precedes the terminator.
When a valid query is received, the CPC 6000 will return {data} terminated by CR and LF.
Floating point data is returned in the current engineering units in exponential format.
7.4.4 Output Formats
Pressure readings are returned in exponential notation in a format according to the OUTFORM com
-
mand as follows:
1. <sp> pressure value <cr><lf>
2. <sp> pressure, units number, mode <cr><lf>
3. <sp> pressure, pressure rate <cr><lf>
4. <sp> pressure, minimum peak, maximum peak<cr><lf>
5. <sp> pressure, active sensor (P or S), active turndown (1-4)<cr><lf>
6. <sp> pressure, control point, “stable” or “slewing”<cr><lf>
7. <sp> pressure, “no barometer” or baro reading <cr><lf>
The space <sp> at the beginning of the string is replaced with an “E” if the CPC 6000 has detected a
remote communication error.
For commands that take boolean data the following strings are acceptable:
0 1
False True
No Yes
Off On