Automated Pressure Calibrator
CPC 6000
52 Mensor/WIKA Operating Instructions - CPC 6000
16 BR? <sp>{value}<cr><lf> Returns the B channel rate.
17
BRS? <sp>{YES or NO}<cr><lf> Returns the B channel rate stable flag.
18 BS? <sp>{YES or NO}<cr><lf> Returns the B channel stable flag.
19 Caldisable {YES or NO} Disables the zero and span commands (default
= YES). When the cal-disable is YES, the zero
and span commands are disabled.
20 Caldisable? <sp>{YES or NO}<cr><lf> Returns whether or not calibration of the active
sensor is disabled.
21
Cerr none Clears the error queue.
22 Chan {A or B} Sets the active channel on the instrument.
23 Chan? <sp>{A or B}<cr><lf> Returns which channel is active.
24 Chanfunc <sp>{Press,peak,rate,baro}<
cr> <lf>
Sets the alternate function mode of the active
channel.
25 Chanfunc? <sp>{string}<cr><lf> Returns the alternate function mode of the
channel.
26
Control none Instrument placed in Control mode.
27 Control? <sp>{YES or NO}<cr><lf> Returns YES if active channel is in Control
mode; NO if otherwise.
28
Crate {Slow, Medium, Fast, Vari-
able} for mid or high range
regulator, {Slow or Fast} for
low range regulator.
Sets the control slew rate for the current chan
-
nel. For Variable rate see: Rsetpt command.
29
Crate? <sp>{string}<cr><lf> Returns the current rate - value corresponds to
the parameters for the CRATE command.
30
Decpt <sp>{4, 5 or 6}<cr><lf> Sets the number of significant digits displayed.
31
Decpt? <sp>{value}<cr><lf> Returns the number of significant digits dis-
played for the active channel.
32 Default none Sets the default values. This command takes
approximately 20 seconds.
33 DIO {0 or 2} 0 turns the digital IO output pin off or low.
2 turns the digital IO output pin on or high.
34 DIO? {value} 0 if input is low and output is low.
1 if input is high and output is low.
2 if input is low and output is high.
3 if input is high and output is high.
35 DOC <mm/dd/yyyy> Sets the date of cal for the active sensor and
turndown.
36
DOC? <sp>{mm/dd/yy}<cr><lf> Returns the date of cal for the active sensor and
turndown.
37 DOM? <sp>{mm/dd/yyyy}<cr><lf> Returns the date of manufacture.
38
Error? <sp>{string}<cr><lf> Returns a description of an error.