Mensor/WIKA Operating Instructions - CPC 6000 57
118 StableWin {%fs value} Sets the stable window as a %FS.
119 StableWin? <sp>{value}<cr><lf> Returns the stable window in % of Span.
120
Standby none Instrument placed in Standby mode, all sole-
noids de-energized.
121
Standby? <sp>{Yes or No}<cr><lf> Returns YES if the active channel is in Standby,
NO if otherwise.
122
Step <sp>{value inside upper and
lower limits and inside the
range of the active sensor
and turndown}<cr><lf>
Sets the control step size for the instrument.
123
Step- {none or value} Jogs the setpoint down one step. If a value is
entered, sets the step size to that value.
124
Step+ {none or value} Jogs the setpoint up one step. If a value is en-
tered, sets the step size to that value.
125
Step? <sp>{value}<cr><lf> Returns the control step for the instrument.
126
Stop none Stop repeating output over serial port. See:
Repeat
127
Units {units code, or output format
text from Measurement Units
Table 11.1 in the Appendix
Section}
Sets the instrument engineering units.
128
Units? <sp>{string}<cr><lf> Returns the instrument units in a text string.
129
Unlock none Releases Acquire locks. See: Acquire? and
Release?
130
UpperLimit {Value inside primary trans-
ducer range}
Sets the upper control limit for the active chan-
nel.
131
UpperLimit? <sp>{value}<cr><lf> Returns the upper control limit for the active
channel.
132
Vent none Instrument placed in Vent mode.
133
Vent? <sp>{Yes or No}<cr><lf> Returns YES if active channel is in Vent mode,
No if otherwise.
134
Versions? <sp>{string}<cr><lf> Returns the versions of the firmware for the
instrument, gpib driver, gate array, installer, and
the graphics library.
135 Volume {value in cc} Sets the system volume in cc’s – only applicable
if the active channel is a low pressure regulator.
136
Volume? <sp>{value}<cr><lf> Returns the current system volume in cc’s – only
applicable if the active channel is a low pressure
regulator.
137
Zero {desired pressure or ?} Sets active sensor and turndown’s zero to en-
tered pressure, or ? to clear previous value. (See
“Caldisable” command above ref# 19)
138 Zero? <sp>{value}<cr><lf> Returns zero offset value for active sensor and
turndown.