7.3.1 Capability Codes
SH1 Full source handshake capability
AH1 Full acceptor handshake capability
T6 Talker with serial poll and unaddress if MLA
L4 Listener with unaddress if MTA
SR1 Full service request capability
RL1 Full remote/local capability including LLO
PP0 No parallel poll capability
DC1 Full device clear capability
DT1 Full device trigger capability
C0 No controller capability
E2 Tri-state outputs
7.3.2 Interface Functions
The CPC 6000 responds to the following IEEE-488 interface functions:
SRQ Service Request: A service request is asserted whenever an error is encountered. When
the bus controller issues a serial poll the error will be cleared. If the host IEEE board in
-
cludes automatic serial polling capability, turn this feature off in order to view all errors (see
ERROR? command).
LLO Local Lockout: The front panel keyboard of the CPC 6000 may be locked by sending LLO
or the command LOCK ON.
GET Group Execute Trigger: When this message is received, the CPC 6000 will save the current
readings until the next time it is addressed as a talker.
GTL Go To Local: A GTL message will cause the CPC 6000 return to local operation and unlock
the keyboard.
DCL Device Clear: When this message is received, the CPC 6000 will clear all errors and buffers
and remain in the REMOTE mode.
SDC Selected Device Clear: The effect is the same as DCL.
EOI End or Identify: May be used as a command or query terminator in the place of, or concur-
rent with, a terminating linefeed.
7.4 RS-232 Serial Communication
The serial communication port allows the CPC 6000 to communicate in RS-232 format with comput
-
ers, terminals, PDAs, or similar hosts.
7.4.1 Cable Requirements
RS-232 communications are transmitted over a three conductor, shielded cable terminated in a stan
-
dard DB9S connector on the instrument end, and usually the same connector on the host end. Figure
7.4.1 illustrates the proper pin-outs. Notice that each pin 2 is connected to pin 3 on the opposite end.
This configuration is commonly referred to as a 9-pin null modem cable.