Automated Pressure Calibrator
CPC 6000
58 Mensor/WIKA Operating Instructions - CPC 6000
7.4.6 CPC 6000 Error Codes
Table 7.4.6 - CPC 6000 Error Codes
Code
Serial
Poll
Byte Description Error String Returned
E00 00h reserved unused code
E01 41h GPIB listen/talk error GPIB LISTEN/TALK ERROR
E02 42h Syntax error UNKNOWN COMMAND
E03 43h Expected a valid _PCS4 command EXPECTED A VALID _PCS4 COMMAND
E04 44h Expected a valid FUNC command EXPECTED A VALID FUNC COMMAND
E05 45h Expected a valid CAL command EXPECTED A VALID CAL COMMAND
E06 46h Expected a valid TEST command EXPECTED A VALID TEST COMMAND
E07 47h Expected a pressure units selection EXPECTED A PRESSURE UNITS SELEC
-
TION OR INVALID TERMINATION STRING
E08 48h Expected a pressure value EXPECTED A PRESSURE VALUE
E09 49h Expected a transducer selection EXPECTED A XDUCER SELECTION
E10 4ah Expected a seconds selection EXPECTED A SECONDS SELECTION
E11 4bh reserved unused code
E12 4ch
reserved unused code
E13 4dh Invalid pressure units selection INVALID PRESSURE UNITS SELECTION
E14 4eh Invalid control pressure value selection INVALID CONTROL PRESSURE VALUE
SELECTION
E15 4fh Invalid rate value selection INVALID RATE VALUE SELECTION
E16 50h Invalid a/d unit selection INVALID A/D UNIT SELECTION
E17 51h Invalid zero offset value selection INVALID ZERO OFFSET VALUE SELECTION
E18 52h Invalid Span offset value selection INVALID SPAN OFFSET VALUE SELECTION
E19 53h Invalid rate units selection INVALID RATE UNITS SELECTION
E20 54h Sensor overrange SENSOR OVERRANGE
E21 55h Sensor underrange SENSOR UNDERRANGE
E22 56h Sensor failure detected SENSOR FAILURE DETECTED
E23 57h Low source pressure LOW SOURCE PRESSURE
E24 58h Regulator failure detected REGULATOR FAILURE DETECTED
E25 59h
reserved unused code
E26 5ah Internal leak detected INTERNAL LEAK DETECTED
E27 5bh Program error detected PROGRAM ERROR DETECTED
E28 5ch Memory error detected MEMORY/COEFFICIENT ERROR DETECT
-
ED
E29 5dh External Leak detected EXTERNAL LEAK DETECTED
E30 5eh High exhaust pressure VACUUM ERROR DETECTED
E31 5fh Transducer error detected XDUCER ERROR DETECTED
Komentarze do niniejszej Instrukcji