Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000 55
75 Peakmax? <sp>{value}<cr><lf> Returns the maximum pressure since peakreset
was sent.
76
Peakmin? <sp>{value}<cr><lf> Returns the minimum pressure since peakreset
was sent.
77 Peakreset none Resets the peak values.
78
Port {value} Sets the Ethernet port of the instrument.
79
Port? <sp>{value}<cr><lf> Returns the Ethernet port of the instrument.
80 Precision none Sets the active channel to the precision mode:
.003% stablewindow and 4 second stabledelay.
81 Ptype {Absolute or Gauge or A or G} Sets the instrument pressure type – non-native
type only works if the optional barometric sen
-
sor is installed.
82
Ptype? <sp>{Absolute
orGauge}<cr><lf>
Returns “Absolute” or “Gauge” for the pressure
type.
83
RangeMax? <sp>{value}<cr><lf> Returns the maximum range of the active trans-
ducer and turndown in the current units.
84
RangeMin? <sp>{value}<cr><lf> Returns the minimum range of the active trans-
ducer and turndown in the current units.
85
Rate? <sp>{value}<cr><lf> Returns the rate reading of the instrument in
current units/second.
86 Rdecpt? <sp>{value}<cr><lf> Returns the number of rate decimal points for
the active channel. See: Resolution
87
Reference {EXTVAC or ATM} Sets the reference type for vacuum pump
connected to the reference, or vented to atmo
-
sphere. This option is only available on native
gauge units.
88 Reference? <sp>{EXTVAC or
ATM}<cr><lf>
Returns the current reference type of the active
channel. See: Reference
89 Release? {15 char string} Returns: <sp>(YES or NO), {string}<cr><lf>
This command is used to release control of the
instrument when it is accessed by more than
one computer. To release control of the instru
-
ment, {string} must be identifier used for the
“Acquire” command. Yes if release is successful.
No if instrument is being controlled with another
computer. Return {string} = identifier of the con
-
trolling computer or AVAILABLE. See: Acquire?
and Unlock.
90 Repeat None Repeats output continually over serial port only.
91 Resolution {4 to 6} Sets the number of significant digits. See: decpt
92
Resolution? <sp>{value}<cr><lf> Returns the number of significant digits.
93
Rsetpt {value} Sets the rate setpoint of the active channel in
current units per second. Must be a value inside
the current sensor range.