3370964.05 04/2014 GB/D/F/E
12
WIKA operating instructions models TR10-L, TC10-L (Ex d)
GB
4. Design and function
4.1 Description
The model TR10-L (resistance thermometer) or TC10-L (thermocouple) electrical
thermometers are comprised of a measuring insert which is built into a certied Ex d
housing. In combination with a ame path tting that is screwed into the head, the measuring
insert functions as a ameproof joint. The measuring insert (TR10-K, TC10-K) is replaceable.
Sensor design in model TR10-L
The measuring resistor is embedded in ceramic powder, heat-resistant potting compound,
cement compound or thermally conductive paste. A one-sided sealed tube, which is welded
to a mineral-insulated cable, forms the external shell of the sensor tip of the measuring
insert.
Sensor design in model TC10-L
The measuring insert of the thermocouple is manufactured from mineral-insulated cable.
The thermocouple consists of the internal leads of the mineral-insulated cable. The weld
spot of the thermocouple is, depending on the design, either ungrounded welded with the
sheath of the mineral-insulated cable or grounded welded.
If the temperature sensor is designed as a grounded thermocouple, the thermocouple is
joined directly to the sheath. Designs with a diameter smaller than 3 mm and with grounded
thermocouples should be considered as galvanically connected with earth potential.
Versions (see gures page 18):
Thermometers without a ame path tting may only be used in combination with a
solid-machined, certied WIKA thermowell with a minimal wall thickness of 1 mm. The
thermometer is marked with IIC and is suitable for use in zone 1.
After installation of a ame path tting into the connection head of the thermometer, a
thermowell is no longer mandatory for certication reasons. In most cases, however,
the use of a thermowell (solid-machined or fabricated, with a wall thickness of 1 mm) is
necessary for process engineering reasons. The thermometer is marked with IIB + H
2
and
is suitable for use in zone 1.
The design of the thermowell can be selected as desired, but the operational process
data (temperature, pressure, density and ow rate) must be taken into account. If a solid
WIKA thermowell is already available or installed, a ame path tting is not necessary.
The thermometers models TR10-L or TC10-L are manufactured by WIKA with certied
Ex d connection heads or connection housings. These enclosures and covers are made
from aluminium or stainless steel. The cover is optionally available with a glass window.