Wika DPT-10 Instrukcja Użytkownika Strona 22

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 96
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 21
à Connect minus side always above the max. level
à The condensate vessel ensures a constant pressure on the minus
side
à For measurements in products with solid content such as e.g. dirty
liquids, the installation of separators and drain valves is
recommended to enable collection and removal of debris and
sediment.
+
min.
max.
1
2
2
3
4
6
5 5
Fig. 18: Measurement setup in closed vessel with superimposed steam
1 Condensate vessel
2 Blocking valves
3 DPT10
4 Precipitator
5 Drain valves
6 3-fold valve block
à Mount DPT10 directly to the vessel
à Connect minus side always above the max. level
à The condensate vessel ensures a constant pressure on the minus
side
à For measurements in products with solid content such as e.g. dirty
liquids, the installation of separators and drain valves is
recommended to enable collection and removal of debris and
sediment.
In closed vessels with
steam layering with uni-
lateral pressure trans-
mitter
22 Dierential pressure transmitter DPT10 - metal measuring diaphragm Probus PA
4 Mounting
37245-EN-100120