Mensor/WIKA Operating Instructions Series CPG 2500 43
Digital Pressure Gauge
SERIES CPG 2500
RWINDOW? <sp>n.nnnnne+nn<cr><lf> Returns rate exponential filter window.
SBAUD 9600, 19200, 38400, 57600 Sets the serial baud rate.
SBAUD? <sp>XXXX<cr><lf> Returns the serial baud data.
SCREENSRC A, B, or A,B, or B,A, or
Delta,A, or Delta,B
Sets the CPG 2500 display mode.
SCREENSRC? <sp>{A,B or A,Delta or
Delta,B}<cr><lf>
Returns screen source as one of the
above options.
SDATA 7 or 8 Sets the serial data bits.
SDATA? <sp>X<cr><lf> Returns the serial data bits number.
SENSORID? <sp>Mensor QRS,SN
XXXXXX,VER V.VV
Returns the active sensor’s serial number
and firmware version.
SPAN desired pressure or ? Sets span on active channel or for ?,
clears previous value, must be >50% FS
and has a 1% limit.
SPAN? <sp>XXXXXXX<cr><lf> Returns span scale factor for active chan-
nel.
SPARITY even, odd, none Sets the serial parity.
SPARITY? <sp>CCCC<cr><lf> Returns the serial parity.
SSTOP 1 or 2 Sets the serial stop bits.
SSTOP? <sp>X<cr><lf> Returns the serial stop bits.
UNITS units code or text in Table
7.1.3.6.
Sets the engineering units for the active
channel.
UNITS? <sp>CCCC<cr><lf> Returns the pressure units of the active
channel in a text string.
UNLOCK none Releases Acquire locks. See ACQUIRE?
and RELEASE?
WINDOW value in current units Sets the exponential filter window for the
active channel.
WINDOW? <sp>n.nnnnne+nn<cr><lf> Returns the exponential filter window for
the active channel.
ZERO desired pressure or ? Sets zero to set pressure or for ?, clears
previous value.
ZERO? <sp>xxxxxxx<cr><lf> Returns zero offset for active channel.