Wika CPG2500 Instrukcja Użytkownika Strona 1

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Urządzenia pomiarowe Wika CPG2500. WIKA CPG2500 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - SERIES CPG 2500

Operating InstructionsDigital Pressure GaugeDigital Pressure Gauge Series CPG 2500 PN 0017851001 FSERIES CPG 2500

Strona 2

10 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25002. Safety Notices2.1 User ResponsibilitiesTo ensure safety,

Strona 3 - Table of Contents

Mensor/WIKA Operating Instructions Series CPG 2500 11Digital Pressure GaugeSERIES CPG 2500iNoticeAdditional safety notices are found throughout this m

Strona 4

12 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500!CautionCAUTION: As with most sensitive electronic equipment

Strona 5

Mensor/WIKA Operating Instructions Series CPG 2500 13Digital Pressure GaugeSERIES CPG 25003. Product Description3.1 General DescriptionThe CPG 2500 Di

Strona 6

14 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25003.3 Front PanelThe CPG 2500 front panel includes a 6.2 inch

Strona 7 - 1. General Information

Mensor/WIKA Operating Instructions Series CPG 2500 15Digital Pressure GaugeSERIES CPG 25003.5 Chassis AssemblyThe chassis assembly acts as the housing

Strona 8

16 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500NOTES

Strona 9

Mensor/WIKA Operating Instructions Series CPG 2500 17Digital Pressure GaugeSERIES CPG 25004. SpecificationsAccuracy specifications presented herein ar

Strona 10 - 2. Safety Notices

18 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500PressuremediaClean, dry, non-corrosive, non-conbustible, non

Strona 11 - Digital Pressure Gauge

Mensor/WIKA Operating Instructions Series CPG 2500 19Digital Pressure GaugeSERIES CPG 25005. Installation5.1 Unpacking the InstrumentIn addition to fu

Strona 12

2 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500Warning!This Warning symbol indicates that danger of injury

Strona 13 - 3. Product Description

20 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25005.4 Rear PanelUp to five pneumatic pressure ports are locate

Strona 14

Mensor/WIKA Operating Instructions Series CPG 2500 21Digital Pressure GaugeSERIES CPG 25005.5.1 Pressure PortConnect a device to be tested to the PRES

Strona 15

22 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500NOTES

Strona 16

Mensor/WIKA Operating Instructions Series CPG 2500 23Digital Pressure GaugeSERIES CPG 25006. Local Operation6.1 General OperationThis section describe

Strona 17 - 4. Specifications

24 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25006.4 Toolbar6.4.1 Contact and Version Information KeyPress th

Strona 18

Mensor/WIKA Operating Instructions Series CPG 2500 25Digital Pressure GaugeSERIES CPG 25006.4.3 Main Screen View KeyPress this key to configure the ch

Strona 19 - 5. Installation

26 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25006.4.4 Remote Setup KeyPress this key to set the operating pa

Strona 20

Mensor/WIKA Operating Instructions Series CPG 2500 27Digital Pressure GaugeSERIES CPG 25006.5.2.1 Display TabPress the display tab to configure the me

Strona 21

28 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500Barometer: If this optional feature was ordered with your CP

Strona 22

Mensor/WIKA Operating Instructions Series CPG 2500 29Digital Pressure GaugeSERIES CPG 25006.5.2.2.2 Delta Pressure Display ModeThis feature is availab

Strona 23 - 6. Local Operation

Mensor/WIKA Operating Instructions Series CPG 2500 3Digital Pressure GaugeSERIES CPG 2500Table of Contents1. General Information 71.1 Warranty 71.2 Im

Strona 24

30 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25006.5.2.2.2.2 Sensor Tab (Delta)The delta function can be sele

Strona 25

Mensor/WIKA Operating Instructions Series CPG 2500 31Digital Pressure GaugeSERIES CPG 25006.5.2.3 Adjust TabThis tab is for adjusting zero and head pr

Strona 26

32 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500iNoticeHead correction parameters are stored separately for

Strona 27

Mensor/WIKA Operating Instructions Series CPG 2500 33Digital Pressure GaugeSERIES CPG 2500Sensor information is displayed showing the serial number, d

Strona 28

34 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25006.5.3.2 Metric Tab6.5.3.3 User Units TabTo enter customized

Strona 29

Mensor/WIKA Operating Instructions Series CPG 2500 35Digital Pressure GaugeSERIES CPG 25007. Remote OperationUse the following screens to set the oper

Strona 30

36 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25007.1.2 Ethernet Tab Press the Ethernet tab to set up the Ethe

Strona 31

Mensor/WIKA Operating Instructions Series CPG 2500 37Digital Pressure GaugeSERIES CPG 25007.1.3 Serial TabPress this tab to set up the serial port par

Strona 32

38 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25007.1.3.2 Command and Query FormatCommands must be sent in ASC

Strona 33

Mensor/WIKA Operating Instructions Series CPG 2500 39Digital Pressure GaugeSERIES CPG 25003. Queries are special instructions in the form: [Command?]

Strona 34

4 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25005.6 Power Up 216. Local Operation 236.1 General Operation 23

Strona 35 - 7. Remote Operation

40 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500A? <sp>n.nnnnne+nn<cr><lf> Returns the A c

Strona 36

Mensor/WIKA Operating Instructions Series CPG 2500 41Digital Pressure GaugeSERIES CPG 2500DOC mm/dd/yyyy Sets the date of cal for the active channel.D

Strona 37

42 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500PEAKMAX? <sp>n.nnnnne+nn<cr><lf> Returns t

Strona 38

Mensor/WIKA Operating Instructions Series CPG 2500 43Digital Pressure GaugeSERIES CPG 2500RWINDOW? <sp>n.nnnnne+nn<cr><lf> Returns r

Strona 39

44 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25007.1.3.6 Units Command Syntax for Measurement UnitsTable 7.1.

Strona 40

Mensor/WIKA Operating Instructions Series CPG 2500 45Digital Pressure GaugeSERIES CPG 250029 pounds per square foot PSF English30 tons per square foot

Strona 41

46 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25007.1.3.8 Remote EmulationThe Mensor DPG 2100 is the previous

Strona 42

Mensor/WIKA Operating Instructions Series CPG 2500 47Digital Pressure GaugeSERIES CPG 2500ADDRESS address sp <value>lfSets the device IEEE-488 a

Strona 43

48 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500DISPLAY? display?lf To determine the current display setting

Strona 44

Mensor/WIKA Operating Instructions Series CPG 2500 49Digital Pressure GaugeSERIES CPG 2500KEYLOCK Keylock sp <value> lfInvoking the KEYLOCK comm

Strona 45

Mensor/WIKA Operating Instructions Series CPG 2500 5Digital Pressure GaugeSERIES CPG 25007. Remote Operation 357.1 Remote Setup 357.1.1 Instrument Tab

Strona 46

50 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500SAVE SAVElf Send a SAVE command to preserve changes made to

Strona 47

Mensor/WIKA Operating Instructions Series CPG 2500 51Digital Pressure GaugeSERIES CPG 2500SPAN SPAN sp <value> lf Use the SPAN command to correc

Strona 48

52 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500UNITS_TABLE not emulatedUNITS_TABLE? UNITS_TABLE? lf not emu

Strona 49

Mensor/WIKA Operating Instructions Series CPG 2500 53Digital Pressure GaugeSERIES CPG 25009 feet of water @ 60°F FTH2O English10 millitorr MTORR Metri

Strona 50

54 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25007.1.3.8.3 DPG II Models 14000/15000 Terse Message Set Emulat

Strona 51

Mensor/WIKA Operating Instructions Series CPG 2500 55Digital Pressure GaugeSERIES CPG 2500The ‘Q’ Command Data/Output Formatn Description Output Forma

Strona 52

56 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500U Command: same as UNITS command.The U command selects the

Strona 53

Mensor/WIKA Operating Instructions Series CPG 2500 57Digital Pressure GaugeSERIES CPG 250025 grams per square centimeter G/CM2Metric26 kilograms per s

Strona 54

58 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25007.1.3.8.4 General SCPI WIKA CommandsSTATus:OPERation :CONDi

Strona 55

Mensor/WIKA Operating Instructions Series CPG 2500 59Digital Pressure GaugeSERIES CPG 2500SENSE[:PRESsure] [z] Selects the sensor to read or modif

Strona 56

6 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25008.2.3 Specifications 668.3 Analog Output 668.3.1 Command Set

Strona 57

60 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500TEST:ELECtronic? Returns electronics status.:RELay<n&g

Strona 58

Mensor/WIKA Operating Instructions Series CPG 2500 61Digital Pressure GaugeSERIES CPG 25007.1.4 IEEE-488 Tab (Option)Press the Remote Setup key and th

Strona 59

62 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25007.1.4.2 Interface FunctionsThe CPG 2500 responds to the foll

Strona 60

Mensor/WIKA Operating Instructions Series CPG 2500 63Digital Pressure GaugeSERIES CPG 25008. Options• Barometric Reference Transducer• Barometer• Anal

Strona 61

64 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500In absolute emulation mode the atmospheric pressure reading

Strona 62

Mensor/WIKA Operating Instructions Series CPG 2500 65Digital Pressure GaugeSERIES CPG 2500Table 8.2.1 - Standard Sea Level Pressure at Various Elevati

Strona 63 - 8. Options

66 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25009000 2743.20 10.50483 21.38811 543.2582010000 3048.00 10.106

Strona 64

Mensor/WIKA Operating Instructions Series CPG 2500 67Digital Pressure GaugeSERIES CPG 25008.3 Analog OutputThe CPG 2500 can be configured at the time

Strona 65

68 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25008.3.1 Command Set for Analog Output OptionCommands can be us

Strona 66

Mensor/WIKA Operating Instructions Series CPG 2500 69Digital Pressure GaugeSERIES CPG 25003. Connect a DMM to the analog output connector on the rear

Strona 67

Mensor/WIKA Operating Instructions Series CPG 2500 7Digital Pressure GaugeSERIES CPG 25001. General Information1.1 WarrantyAll products manufactured b

Strona 68

70 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500Uncertainties: 0-1 VDC 0.010% FS 0-5 VDC 0.010% FS 0-1

Strona 69

Mensor/WIKA Operating Instructions Series CPG 2500 71Digital Pressure GaugeSERIES CPG 25009. MaintenanceThe CPG 2500 was designed for maintenance-free

Strona 70

72 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25009.2 Spare PartsTable 9.2 lists the spare parts for the CPG 2

Strona 71 - 9. Maintenance

Mensor/WIKA Operating Instructions Series CPG 2500 73Digital Pressure GaugeSERIES CPG 250010. CalibrationThe CPG 2500 automatically adjusts the pressu

Strona 72

74 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 250010.4 SetupThe following illustration for calibration setup s

Strona 73 - 10. Calibration

Mensor/WIKA Operating Instructions Series CPG 2500 75Digital Pressure GaugeSERIES CPG 250010.5 Calibration The calibration screen is password protecte

Strona 74

76 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 250010.5.1 Sensor TabPress the Sensor tab to view information ab

Strona 75

Mensor/WIKA Operating Instructions Series CPG 2500 77Digital Pressure GaugeSERIES CPG 250010.5.2.2 Restore Factory CalPress OK to restore the factory

Strona 76

78 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 250010.5.3 Edit TabPress the Edit tab to adjust the calibration

Strona 77

Mensor/WIKA Operating Instructions Series CPG 2500 79Digital Pressure GaugeSERIES CPG 250010.5.4.1 Procedure 1: For ‘as found’ data1. Press the Calib

Strona 78

8 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 25001.3 FCC Radio Frequency Emission NoticeThis equipment has be

Strona 79

80 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 2500NOTES

Strona 80

Mensor/WIKA Operating Instructions Series CPG 2500 81Digital Pressure GaugeSERIES CPG 250011. AppendixTable 11.1 - Measurement Units (unitno)Code Desc

Strona 81 - 11. Appendix

82 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 250029 pounds per square foot PSF30 tons per square foot TSF31 p

Strona 82

Mensor/WIKA Operating Instructions Series CPG 2500 83Digital Pressure GaugeSERIES CPG 250011 inSW 0°C 3.5% salinity 26.92334 0.0371425012 ftSW 0°C 3.5

Strona 83

84 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 250011.3 Conversion Factors, MillitorrThe following table lists

Strona 84

Mensor/WIKA Operating Instructions Series CPG 2500 85Digital Pressure GaugeSERIES CPG 250027 MSW 0°C 3.5% salinity 0.00001322347 75623.1166328 OSI 0.0

Strona 85

86 Mensor/WIKA Operating Instructions Series CPG 2500Digital Pressure GaugeSERIES CPG 250015 mBar 1.00000E-02 1.00000E+0216 mmH2O 4°C 1.019744E-01 9.8

Strona 87

Mensor Corporation201 Barnes DriveSan Marcos, Texas 78666Phone: 512.396.4200Fax: 512.396.1820Web site: www.mensor.comemail: [email protected] 2500 O

Strona 88

Mensor/WIKA Operating Instructions Series CPG 2500 9Digital Pressure GaugeSERIES CPG 2500or incidental damages including damages for loss of business

Komentarze do niniejszej Instrukcji

Brak uwag