Mensor/WIKA Operating Instructions Series CPG 2500 41
Digital Pressure Gauge
SERIES CPG 2500
DOC mm/dd/yyyy Sets the date of cal for the active channel.
DOC? <sp>mmddyy<cr><lf> Returns the date of cal for the active chan-
nel.
DOM? <sp>mm/dd/yyyy<cr><lf> Returns the date of manufacture.
FILTER Off, Low, Normal, High Sets the reading filter 0, 80%, 92%, 95%.
FILTER? <sp> (filter)<cr><lf> Returns the reading filter.
GASDENSITY Value in lb/cuft Sets the head pressure gas density.
GASDENSITY? <sp>n.nnnnne+nn<cr><lf> Gets the head pressure gas density.
GASTEMP Value in degrees F Sets the head pressure gas temperature.
GASTEMP? <sp>n.nnnnne+nn<cr><lf> Gets the head pressure gas temperature.
GATEWAY nnn.nnn.nnn.nnn Sets the Ethernet gateway address.
GATEWAY? <sp>nnn.nnn.nnn.nnn
<cr><lf>
Gets the Ethernet gateway address.
HEIGHT Value in inches Sets the head pressure height.
HEIGHT? <sp>n.nnnnne+nn<cr><lf> Gets the head pressure height.
ID? <sp> MENSOR, 2500,
ssssss,v.v.vv
Ssssss is the serial number, v.v.vv is the
CPG 2500 software version.
IP nnn.nnn.nnn.nnn Sets the IP address of the instrument.
IP? <sp>nnn.nnn.nnn.nnn
<cr><lf>
Returns the IP address of the instrument.
KEYLOCK yes or no Locks or unlocks keyboard.
KEYLOCK? <sp>(yes or no)<cr><lf> Returns yes or no.
LISTRANGE? PRI,1, min,max Returns the range of the installed sensor
for the active channel.
LOCALGRAVITY value in ft/s^2 Sets the local gravity in feet/sec^2.
LOCALGRAVITY? <sp>n.nnnnne+nn<cr><lf> Returns the local gravity in feet/sec^2.
NETMASK nn.nnn.nnn.nnn Sets the Ethernet network mask.
NETMASK? <sp>nnn.nnn.nnn.nnn
<cr><lf>
Gets the Ethernet network mask.
OUTFORM 1 to 4 - (see Table 7.1.3.6) Sets the output format.
OUTFORM? <sp>X<cr><lf> Returns the output format (see 7.1.3.4).