WIKA operating instructions model TR25 (Ex i)
GB
7.4.2 Possible installation methods with the marking II 1/2 Ex ib IIC T6 Ga/Gb or
II 1/2 D Ex ib IIIC T65 °C Da/Db
The sensor or thermowell tip protrudes into Zone 0. The housing or connection head is in Zone 1
(Zone 21) or Zone 2 (Zone 22). It is sucient to use an Ex ib type circuit.
Zone separation is guaranteed if suciently-tight (IP 67) process connections are used.
Examples of suitable process connections include gas-tight standardised industrial anges,
threaded connections or pipe connections.
The welded parts, process connections, compression ttings, thermowells or housings used
must be designed such that they withstand all inuencing variables resulting from the process,
such as temperature, ow forces, pressure, corrosion, vibration and impacts.
Zones 1, 2 or
Zones 21, 22
Hazardous area
Safe area
Thermowell
Thermowell
welded
TWxx
TWxx
Process
connection
Connection head/
Field housing
Tx10-B
Tx10-C
Tx10-C
Tx10-D
Tx10-H
Tx10-H
Option:
with built-in
transmitter:
e.g. T32
Tx10-A
Process
connection
Compression
tting
Associated
electrical equipment
Intrinsically
safe supply or
suitable barrier
Intrinsically
safe supply or
suitable barrier
Intrinsically
safe supply or
suitable barrier
Intrinsically
safe supply or
suitable barrier
Tx10-A
Tx10-A
Tx10-A
Zones 0, 1, 2 or
Zones 20, 21, 22
7. Information on mounting and operation in hazardous areas