6 Commissioning, operation
66 WIKA Operating Instruction, Pascal ET and Pascal ET/IS
GB
Press SETUP to set the ramp:
Figure 72 Ramp parameters setup
Here above a ramp has been set: starting from 0 V, in 11 steps it reaches 1 V after 20 sec.: then it
remains standstill for 5 sec. The ramp will be executed 2 times (Cycles num.).
Press OUT A to start the ramp function: the instrument will show the main screen and the writing
RMP will be displayed to show that the function ramp is active.
Figure 73 - Ramp running
If another channel is assigned to OUT B, here below you can see how the Ramp will be shown:
Figure 74 Ramp setup with both OUTs
OUT A & B are both available. The Ramp can start for both the channels simultaneously