6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET and Pascal ET/IS 19
6.1.2.3 Pressure module
It is possible to install two pressure modules, each one can support one external transducer.
Accordingly 2 external sensors can be connected at the same time.
Transducers selection with range and their resolutions are in the following table.
External transducers *
* Other ranges available on request
6.1.2.4 HART module
The HART module allows communication with HART instruments (usually transmitters) permitting
to acquire digital measurements, instrument data and modifying settings. It can be inserted in the
front panel in place of OUT A or OUT B module.
It is a fully galvanically isolated board (like IN and OUT modules) from internal electronics of the
Pascal.
The module allow the 24V to supply directly the output loop for the transmitter.
Further, it can supply a 250Ω load resistance electronically switchable necessary to HART
communication, permitting to avoid the use of an external one.
The following figure shows the pin connections in HART module:
Figure 3 – HART module