6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET and Pascal ET/IS 45
GB
The instrument requires the selection of the role:
Figure 45 - Channel role selection
NONE to use the channel for simple measurement
REF to use the channel as reference in a comparison calibration
DUT to use the channel as measurement of a Device Under Test in a comparison
calibration
Press one of the above button (NONE, REF, DUT) to define the channel role.
Pressing REF or DUT the procedure continues as shown below:
I/O to assign a particular I/O module to the software channel.
KEYBOARD to setup by keyboard a particular value read from an external device.
Figure 46 - Channel mode selection
Then press I/O to assign a particular I/O module to the software channel, or press keyboard to
setup a particular value read from an external device.