6. Commissioning, operation
Measuring assemblies
The preferred measuring assemblies for various possible applications are
specied in DIN 19216.
The following schematic diagram shows a recommended assembly for liquid
media. As throttling devices, dierential pressure transducers should be provided
in accordance with DIN 1952 (issue 07.82), now replaced by EN 5167/1.
The dierential pressure lines must be made from metal, their bore must not
be less than 4 mm and the eective length between the valve manifold and the
dierential pressure gauge must be at least 500 mm.
In addition, the length and the bore of
the pressure lines should be such that,
with cold lines, the response time of
the pressure gauge is not more than 5
seconds.
The connections of the dierential
pressure lines must be welded, brazed or
screwed using metal sealing elements.
Shut-o valves in dierential pressure
lines must only be operated using tools.
Dierential pressure
transducer
(throttling device in the
process line)
Dierential pressure
instrument
(Dierential pressure
gauge DELTA-comb)
Shut-o valves
(valve manifold)
Dierential pressure
line
6.2 Electrical connection
■
The electrical connection must only be made by qualied skilled personnel.
■
Connection details and switching functions are given on the product label.
Connection terminals and earth terminal are appropriately marked.
■
The mains connection lines to be provided must be dimensioned for maximum
instrument power consumption and comply with IEC 227 or IEC 245.
■
The instruments must be connected to the equipotential bonding of the plant.
Performance data (see „Specications“)