WIKA operating instructions models PGS10, PGS11 and PGS21
8
Models PGS10, PGS11 and PGS21
Pressure limitation
- Steady 3/4 x full scale value
- Fluctuating 2/3 x full scale value
- Short time Full scale value
Temperature eect When the temperature of the measuring system
deviates from the reference temperature (+20 °C):
max. ±0.4 %/10 K of the span
Ingress protection
- Enclosing case per EN 60529 / IEC 529
Model PGS10 and PGS11: IP 41
(for PGS11.040 with VdS: IP 54)
Model PGS21: IP 65
Operating temperature
- Ambient -20 ... +60 °C
- Medium Maximum +60 °C
For further specications see data sheet PV 20.01, PV 21.01, PV 21.02,
SP 21.01, SP 21.03 and the order documentation.
3. Specications
3. Specications