WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
GB
7. Maintenance and cleaning / 8. Faults
7. Maintenance and cleaning
7.1 Maintenance
Repairs must only be carried out by the manufacturer.
7.2 Cleaning
CAUTION!
Clean the instrument with a moist cloth.
Electrical connections must not come into contact with moisture.
Residual media on the calibration system can result in a risk to
persons, environment and equipment. Take sucient precautionary
measures.
For information on returning the instrument see chapter 9.2 "Return".
7.3 Recalibration
DKD/DAkkS certicate - Ocial certicates:
We recommend that the calibration system is regularly recalibrated by the manufacturer,
with time intervals of approx. 12 months. Every factory recalibration includes, addition-
ally, an extensive free-of-charge check of all system parameters with respect to their
compliance with the specication. The basic settings will be corrected if necessary.
8. Faults
8.1 Digital pressure gauge fault
Display indication Causes Measures
Low battery voltage
functioning is only guaranteed for a
short period of time
Replace alkaline batteries
OL
-OL
Reading above or below the
measuring range
Adjust the line pressure to the
admissible range
No display or undenable
characters
Battery at
Power supply unit is set incorrectly
or the polarity is revered
System error
Instrument faulty
Replace alkaline batteries
Set the power supply unit
correctly and ensure correct
polarity
Switch o the instrument, wait
for a short period of time and
switch on again
Send to the manufacturer for
repair.