6. Commissioning, operation
Compare and record the displayed value of the test
item with the displayed value of the digital pressure
gauge.
Open the pressure relief valve after calibration to
depressurize the system.
The test item can be separated from the calibration
system.
To improve the readability of the display, the backlighting can be activated
by means of the ENTER key.
The calibration system can be dismount now in the
reverse order.
WARNING!
Only disconnect once the system has been depressurised!