Wika T12 Instrukcja Użytkownika Strona 16

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 72
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 15
GB
3211517.0707/2010GB/D
16 WIKAoperatinginstructionstemperaturetransmitterT12
Mounting example:
3173801.A
Mounting on the measuring insert
Mountthetransmitteronthecircularplateofthemeasuringinsert
usingtwocountersunkM3screwsperDINENISO2009.Appropriate
threadedinsertshavebeenpress-ttedintheundersideofthecase.
Assumingthecountersinkingiscarriedoutcorrectly,thepermissible
screwlengthcanbecalculatedasfollows:
l
max.
=s+4mm
with
l
max.
 Lengthofscrewinmm
s Thicknessofcircular
plateinmm
3173801.B
max.4
Circularplate
l
s
6. Commissioning, operation
6. Commissioning, operation
Inhazardousareas,onlyusetemperaturetransmitters
thatareapprovedforthosehazardousareas.The
approvalismarkedontheproductlabel.
6.1 Mounting
6.1.1 Transmitter in head mounting version (model T12.10)
Thetransmittersforheadmountingaredesignedtobemounted
onameasuringinsertwithinaFormB,DINconnectionhead,with
extendedmountingspace.Theconnectionwiresofthemeasuring
insertmustbeapprox.50mmlongandinsulated.
Przeglądanie stron 15