40 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
13. Heating and cooling times
13.5 Dry-well temperature calibrator, model CTD9100-650
Measuring conditions
■
All times refer to a Pt100 reference sensor with Ø 6 mm
■
The reference sensor is positioned, at full depth, centrally
in the insert
■
All times are transition times and do not account for the
required settling time
■
The measurements were carried out at a room tempera-
ture of approx. 23 °C
Insert
Heating up: Time Cooling: Time
25 °C to 40 °C 0:54 min 650 °C to 600 °C 2:25 min
40 °C to 50 °C 0:22 min 600 °C to 550 °C 2:33 min
50 °C to 100 °C 1:18 min 550 °C to 500 °C 2:55 min
100 °C to 150 °C 1:06 min 500 °C to 450 °C 3:27 min
150 °C to 200 °C 1:03 min 450 °C to 400 °C 4:01 min
200 °C to 250 °C 1:05 min 400 °C to 350 °C 4:39 min
250 °C to 300 °C 1:06 min 350 °C to 300 °C 5:36 min
300 °C to 350 °C 1:09 min 300 °C to 250 °C 6:46 min
350 °C to 400 °C 1:21 min 250 °C to 200 °C 8:32 min
400 °C to 450 °C 1:30 min 200 °C to 150 °C 11:22 min
450 °C to 500 °C 1:32 min 150 °C to 100 °C 17:01 min
500 °C to 550 °C 1:38 min 100 °C to 50 °C 52:37 min
550 °C to 600 °C 1:55 min 50 °C to 40 °C 15:23 min
600 °C to 650 °C 2:33 min 40 °C to 25 °C 1:01:58 min