3. Specications
3. Specications
Measuring range
Pressure
Relative pressure -0.8 ... +2 bar -0.8 ... +10 bar -0.8 ... +20 bar
Overpressure safety 4 bar 13 bar 40 bar
Burst pressure 20 bar 20 bar 135 bar
Absolute pressure 0 ... 2 bar abs. 0 ... 10 bar abs.
Overpressure safety 4 bar abs. 13 bar abs.
Burst pressure 20 bar abs. 20 bar abs.
Accuracy 0.025 % FS
Resolution 5-digit
Current
Measuring range 0 … 24 mA (max. load 1,000 Ω)
Resolution 1 µA
Accuracy 0.015 % of reading ±2 µA (simulation and measurement)
Voltage
Measuring range DC 0 … 30 V
Resolution 1 mV
Accuracy 0.015 % of reading ±2 mV (measurement)
Temperature
Measuring range -40 … +150 °C
Resolution 0.01 °C
Accuracy 0.015 % of reading ±20 mΩ, or 0.2 °C for complete measuring chain
(Pt100 resistance thermometer and CPH6600)