41WIKA operating instructions, model CPH6600
14005103.03 11/2012 GB/D
GB
6. Commissioning, operation
6. Use the arrow keys to set the upper limit of the measuring range; select
"DONE SET" when nished.
7. Use the arrow keys to set lower limit of the measuring range, and select
"DONE SET" when nished. The "%-ERROR" mode will be ready to use.
The lower and upper limit of the measuring range will be saved in
non-volatile memory until they are changed again by the user for the
internal pressure sensor and the external pressure modules.