Automated Pressure Calibrator
CPC 6000
92 Mensor/WIKA Operating Instructions - CPC 6000
3. Enter the current password. As each number is pressed an ‘*’ appears in the ‘Password’ window.
When completed touch [OK] to unlock the calibration data screen.
4. The Calibration Data screen of Figure 10.6.1C is displayed. Use this screen to change the password
to perform functions for 1 point calibration, 2 point calibration, restore factory cal, and head correc
-
tion. Touch [Change Password] and the Change Password window appears as in Figure 10.6.1D.
Figure 10.6.1C - Calibration Data Screen (unlocked)
Figure 10.6.1D - “Change Password” Window
5. At the Change Password screen enter from one to six digits for a new password. As each number
is pressed it appears in the Password window.
6. Before proceeding, review the displayed digits for accuracy. A mistake here could prevent future
access to this screen. To make a correction use [CE] to backspace through the entries and then im
-
mediately re-enter the correct numbers.
CAUTION: The password is seldom used and is easily forgotten. After a change
write down and save the new number. If the password is lost, contact Mensor.
!