6. Commissioning, operation
Press ENTER/MENU to conrm the required setting. Si_ appears on the display. Use
the arrow keys ▲▼ to select the sign:
Si_ = the number to be entered is in the negative range (below 0.00 °C)
Si
┘
= the number to be entered is in the positive range
Press ENTER/MENU to conrm the required setting. Fd 0 appears on the display.
Select the range using the arrow keys ▲▼:
Fd 0 = lower than 1,000 °C
Fd 1 = higher than 1,000 °C
Press ENTER/MENU to conrm the required setting. 00.00 appears on the display (or
similar). Enter the temperature of your reference now. Change the digits using the arrow
key ▲. Use the arrow key ▼ to switch to the next digit. After completing the input of the
temperature conrm by pressing ENTER/MENU.
Press [ESC] to return to the measuring mode.
The physical calibration oP2 cannot be cancelled by means of the ESC
key. If required, the calibration can be cancelled by switching o the
measuring instrument.