14 WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio, WU-2X
11553678.02 GB/D/F/E 10/2012
Specications Models WU-20, WU-25 and WU-26
WU-20 / WU-25
WU-26
Pressure ranges psi 14.5 25 60 100 160 250 350 500 1000 1500
bar 1 1.7 4 7 11 17 25 36 70 100
Over pressure safety *
)
psi 120 120 120 210 320 500 750 1100 2100 3000
Burst pressure *
)
psi 1800 1800 1800 2200 2600 4800 6200 5800 8000 10500
Pressure ranges psi 2000 3000 5000
bar 145 225 360
Over pressure safety *
)
psi 4200 6600 10500
Burst pressure *
)
psi 10500 10500 12000
Other pressure ranges, pressure units (e.g. MPa, kg/cm
2
) and absolute pressure on request
*
)
1 psi = 0.069 bar
Measuring principle Metal thin-lm sensor
Materials
■
Wetted parts
- Pressure Connection 316L VIM/VAR
- Pressure sensor 2.4711 / UNSR 30003
■
Case 304 SS
Particle test ≤ 0.1 µm Particle 0.1 ptc / ft³ according to Semi E49.8
Inboard helium leak test < 1 x 10
-9
mbar l/sec (atm STD cc/sec) according to Semi F1
Surface nish Electropolished, typical Ra ≤ 0.13 µm (RA 5); max. Ra ≤ 0.18 µm (RA 7) exceeds Semi F19
Dead volume cm
3
WU-20 < 1.5, WU-25 < 1, WU-26 < 1
Permissible Medium Special gas / Vapour / Liquid
Power supply U+ U+ in VDC 10 ... 31 with output signal 4 ... 20 mA / 0 ... 5 V
14 ... 31 with output signal 0 ... 10 V
Signal output and RA in Ohm 4 … 20 mA, 2-wire RA
≤ (U+ – 10 V) / 0.02 A
maximum ohmic load RA 0 ... 5 V, 3-wire RA > 5 k
0 ... 10 V, 3-wire RA > 10 k
Power Pi W 1
Adjustability zero % of span -3.5 up to +3.5 (via potentiometer) Current output signal
-2 up to +3.5 (via potentiometer) Voltage output signal
7. Starting, operation