11WIKA operating instructions pressure transmitter, model S-20
14043170.02 07/2014 GB/D/F/E
GB
3. Specications
Specications
Continuous vibration resistance (per
IEC 68-2-6)
10 g
Shock resistance (per IEC 68-2-27) 100 g, 6 ms (500 g, 1 ms for heavy-duty connector)
Service life 100 million load cycles (10 million load cycles for measuring ranges > 600 bar/7,500 psi)
Free-fall test (following IEC 60721-3-2)
■
Individual packaging 1.5 m (5 ft)
■
Multiple packaging 0.5 m (1.6 ft)
■
PE bag 0.5 m (1.6 ft)
Electrical protective measures The electrical protective measures are not valid for ratiometric output signals.
■
Short-circuit resistance S+ vs. U-
■
Reverse polarity protection U+ vs. U-
■
Resistance to overvoltage DC 40 V
■
Insulation voltage DC 750 V
Materials of wetted parts
■
Relative measuring ranges Measuring ranges ≤ 10 bar (150 psi): 316L
Measuring ranges > 10 bar (150 psi): 316L + 13-8 PH
■
Absolute measuring ranges Measuring ranges ≤ 1,000 bar (10,000 psi): ASTM 630 and 13-8 PH
Measuring ranges > 1,000 bar (10,000 psi): 316L + 13-8 PH
Materials of non-wetted parts
■
Case 316 Ti
■
Zero point adjustment ring PBT/PET GF30
■
Angular connector DIN 175301-803 A PBT/PET GF30
■
Angular connector DIN 175301-803 C PBT/PET GF30
■
Circular connector M12 x 1 (4-pin) PBT/PET GF30
■
Circular connector M12 x 1 (4-pin,
metallic)
316L
■
Bayonet connector (6-pin) 316L + Al
■
Field case 316L, 316Ti