2266939.05 D/GB 06/2006
Universal IS Pressure Transmitter UniTrans
Technical Data
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany
WIKA A
lexand
er Wieg
and Gmb
H & Co. KG · Al
exand
er-Wie
gand-
Str. · 6391
1 Kling
enbe
rg ·
(09372) 132 - 710 · Fax - 706 · E-mail: sup[email protected] · www.wika.de104
6.3 Construction
Overall deviation
(at +10 °C ... +40 °C)
better than 0.15 % of span for pressure
ranges of <
1000 bar
better than 0.6 % for pressure ranges of
> 1000 bar
Load R
A
< (U
B
–12 V)/0.023 A
with R
A
in Ohm and U
B
in Volt
Fault signal 3.6 mA or 21 mA, programmable
Integration time 0 s, 1 s, 5 s, 20 s, 40 s, programmable
Adjustment of the span Up to Turn Down 1 : 20
Integrated lightning protection optional
Zero point adjustment -2.5 ... 99 %
Process connections
Model IUT-10
Model IUT-11
Model IUT-11 EHEDG version
G 1/2 per EN 837 (max. 1600 bar)
1/2 NPT per "nominal size for US
standard tapered pipe
thread NPT"
(max. 1600 bar)
M 16 x 1,5 with sealing cone
(>
1600 bar)
M 20 x 1,5 with sealing cone
(>
1600 bar)
9/16 - 18 UNF innen F 250-C
(>
1600 bar)
G 1 flush diaphragm with o-ring
(0 ... 0.4 up to 0 ... 1.6 bar)
G 1/2 flush diaphragm with o-ring
(0 ... 6 bis 0 ... 600 bar)
G 1 1/2 flush diaphragm with o-ring
(0 ... 0.4 bis 0 ... 16 bar)
G 1 flush diaphragm with o-ring
(0...0.4 bis 0...16 bar)
Komentarze do niniejszej Instrukcji