27WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
7. Interface
Command Answer function
Control mode
:pa <-110 ... 110> Increase the set pressure demand by x percent
-110 ... +110 --> change in set pressure demand in %
:pd Decrease set pressure demand by currently-set step size in %
(Step DOWN).
:pr <-1100 ... 11000> Adjust the current working and measuring ranges
-1,100 ... +11,000 --> new measuring range in 0.01 % FS
:ps <-10 ... 110> Percentage set pressure demand -10 ... +110 --> set pressure
demand in %
:pu Increase set pressure demand by currently-set step size in %
(Step UP).
:saz<0 | 1> Set auto zero (in measuring and control modes)
0 --> o
1 --> on
:sbr<0 ... 1> Set baud rate RS-232
0 --> 1200
1 --> 2400
2 --> 4800
3 --> 9600
4 --> 14400
5 --> 19200
6 --> 28800
7 --> 38400
8 --> 56000
9 --> 57600
:sbu<0 ... 1> Set baud rate USB
0 --> 1200
1 --> 2400
2 --> 4800
3 --> 9600
4 --> 14400
5 --> 19200
6 --> 28800
7 --> 38400
8 --> 56000
9 --> 57600
:sci<n | u | r> Select active interface
Set communication interface
n --> Interface o
u --> USB active
r --> RS-232 active
sdb <0 ... 100> Display brightness
0 ... 100 --> brightness in %
:spu <0 ... 9> Pressure unit
0 --> Pa
1 --> hPa
2 --> ka
3 --> mbar
4 --> bar
5 --> Torr
6 --> mmHg
7 --> inHg
8 --> psi
9 --> mmH
2
O
10 --> inH
2
O
:ssl <d | e> System language
d = German
e = English