WIKA operating instructions pressure gauge model 6 per ATEX
6. Commissioning, operation
Installation
Nominal position per EN 837-3 / 9.6.7 Figure 9: 90° (
)
Process connection: lower mount (LM) or model 63X.50 back mount (BM)
After mounting, set the compensating valve (if available)
from CLOSE to OPEN.
In order to avoid any additional heating, the instruments must not be exposed
to direct solar irradiation while in operation!
Admissible ambient and working temperatures
When mounting the pressure gauge it must be ensured that, taking into
consideration the inuence of convection and heat radiation, no deviation above
or below the permissible ambient and medium temperatures can occur. The
inuence of temperature on the display accuracy must be observed.
Permissible vibration load at the installation site
The instruments should always be installed in locations free from vibration.
If necessary, it is possible to isolate the instrument from the mounting point
by installing a exible connection line between the measuring point and the
pressure gauge and mounting the instrument on a suitable bracket.
Liquid-lled gauges (model 633.50) may only be used within the following
limits:
Frequency range < 150 Hz and
Acceleration < 0.5 g (5 m/s
2
)
The liquid lling must be checked on a regular basis.
The liquid level must not drop below 75 % of the gauge diameter.
Commissioning
During the commissioning process pressure surges must be avoided at all costs.
Open the shut-o valves slowly.