INSTRUCTION MANUAL
NI-492E
REV. 0 01/05
TEMPERATURE SWITCHES SERIES TXS,TXN,TXA
All data, statements and recommendations supplied with this manual are based on information believed by us to be reliable. As the
conditions of effective use are beyond our control, our products are sold under the condition that the user himself evaluates such
conditions before following our recommendations for the purpose or use foreseen by him.
The present document is the property of ETTORE CELLA SPA and may not be reproduced in any form, nor used for any purpose
other than for which it is supplied.
WEATHERPROOF (Series TXS) AND FLAMEPROOF (Series TXA, TXN)
INSTRUMENT STANDARD INSTRUMENT WITH JUNCTION BOX
Stems Type C-Q-R Stem Type B
Stem Type B: Y =125mm ; X = 50mm Stem Type Q: K = 5m ; Y max.=900mm
Stem Type C: K = 2m ; Y max.=350mm Stem Type R: K = 10m ;Y max.=1800mm
WEIGHT 0,4kg Dimensions in mm WEIGHT 1kg Dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings.
CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given
in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
INSTALLATION IS TO BE CARRIED OUT ONLY AFTER CHECKING THAT INSTRUMENT CHARACTERISTICS ARE
CONSISTENT WITH PROCESS AND PLANT REQUIREMENTS.
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
CONTENTS:
1 - GENERAL
2 - OPERATING PRINCIPLE
3 - IDENTIFICATION PLATE AND MARKINGS
4 - SPECIAL CONDITIONS FOR SAFETY USE (X)
5 - SET POINT ADJUSTMENT
6 - SET POINT CALIBRATION
7 - INSTRUMENT PLUMBING
8 - MOUNTING AND CONNECTIONS
9 - PUTTING INTO OPERATION
10 - FUNCTIONAL VERIFICATION
11 - TROUBLE SHOOTING
12 - STOPPING AND DISMOUNTING
13 - DEMOLITION
14 - NOTE FOR INSTALLATION
SAFETY INSTRUCTIONS FOR USE IN EXPLOSIVE ATMOSPHERES
RECOMMENDATIONS FOR THE SAFE USE OF THE INSTRUMENT
DOCUMENT CORRELATED
to document authenticated with certificates
N° CESI 05 ATEX 002X
N° CESI 05 ATEX 003
Przeglądanie stron 0
1 2 3 4 5 6

Podsumowanie treści

Strona 1 - NI-492E

INSTRUCTION MANUAL NI-492E REV. 0 01/05 TEMPERATURE SWITCHES SERIES TXS,TXN,TXA All data, statements and recommendations supplied with this manu

Strona 2

INSTRUCTION MANUAL NI-492E REV. 5 11/02 1 - GENERAL 1.1 FOREWORD The wrong choice of a series or a model, as well as the incorrect installatio

Strona 3

INSTRUCTION MANUAL NI-492E Rev. 0 01/05 6.1 PRELIMINARY OPERATIONS Slide up the adjustment cover (Fig. 2, 1). 6.2 CALIBRATION CIRCUIT AND OPERAT

Strona 4 - DO NOT TURN

INSTRUCTION MANUAL NI-492E Rev. 0 01/05 Fig. 6 - Mounting of ground device DO NOT TURN 8.3.2 MOUNTING WITH CABLE PROTECTED BY FLEXIBLE TUBI

Strona 5

INSTRUCTION MANUAL NI-492E Rev. 0 01/05 9 - PUTTING INTO OPERATION As the signal transmitted by the instrument is used in a complex system, it i

Strona 6 - Rev. 0 01/05

INSTRUCTION MANUAL NI-492E Rev. 0 01/05 ETTORE CELLA SPA Viale de Gasperi, 48 - Casella Postale (P.O. Box) 58 - I 20010 Bareggio (MILANO) ITALY T